Personal tools
You are here: The Britz-McKibbin Laboratory > Publications > Quantitative assay for epinephrine in dental anesthetic solutions by capillary electrophoresis.

P Britz-Mckibbin, A R Kranack, A Paprica, and D D Chen (1998)

Quantitative assay for epinephrine in dental anesthetic solutions by capillary electrophoresis.

Analyst, 123(7):1461-3.

A simple and robust method for the separation and quantification of epinephrine in dental anesthetic solutions was developed. The method allows the direct injection of high salt solutions without sample pre-treatment. Large sample plugs (5.7% of the total capillary length) are used for epinephrine determination by selective analyte focusing in capillary electrophoresis. The concentration detection limit for epinephrine is about 5.0 x 10(-7) M (90 ng ml-1) with a commercial UV detector. The separation protocol was validated in terms of its precision, linearity, accuracy and specificity.

automatic medline import
 

Document Actions